labsolutions.py use "R.Time (min)" as start marker. But the .txt file may have pump,oven,etc. data. Some like
[LC Status Trace(Pump B Degasser Pressure)] Interval(msec),1000 # of Points,361 Start Time(min),0.000 End Time(min),6.000 Intensity Units,MPa Intensity Multiplier,0.0980665 R.Time (min),Intensity 0.00000,0.00 0.01667,0.00 0.03333,0.00 0.05000,0.00
I think using "[PDA 3D" as data section marker is a solution. FYI: Here is my workaround code. https://gist.github.com/akirayou/16362e28f946dd0c75b21c9933329f15
The text was updated successfully, but these errors were encountered:
No branches or pull requests
labsolutions.py use "R.Time (min)" as start marker. But the .txt file may have pump,oven,etc. data.
Some like
I think using "[PDA 3D" as data section marker is a solution.
FYI: Here is my workaround code.
https://gist.github.com/akirayou/16362e28f946dd0c75b21c9933329f15
The text was updated successfully, but these errors were encountered: