This is derived from parent issue #166 and a breakdown the issues related to node INGV.
These are stations, for which I had downloaded data in recent years, but the metadata is not available now
IV.T0110
Here it looks like the PaZ are agiven in HERTZ, but the units are given as RADIANS – this issue has already been discussed for IV.VINO, see #68. It affects the Guralp sensor epochs.
This is derived from parent issue #166 and a breakdown the issues related to node INGV.
These are stations, for which I had downloaded data in recent years, but the metadata is not available now
Here it looks like the PaZ are agiven in HERTZ, but the units are given as RADIANS – this issue has already been discussed for IV.VINO, see #68. It affects the Guralp sensor epochs.
AC network
no data epochs?
epochs open since 2010, but there were some malfunctions and hence the epochs could be closed
Here the instrumentation was upgraded by temporary BB deployment, but epochs are not
missing station