To install a mini inst, you need to install the EPICS backend and add the following in the \NDXxxxxx\configurations\ area.
-
startup.txt: contains a list of the IOCs to start (use the IOC name as in the IOCs list) -
globals.txt: add macro values you want the IOCs instartup.txtto start up with. To do this, add a line per macro of the format<IOC_name>__<macro_name>=<macro_value>
You can start up the mini inst using the normal start_ibex_server and stop it with stop_ibex_server. It will look for the startup.txt and start up the mini inst instead of the full IBEX server if it exists.
If you update a configuration in globals.txt then don't forget to restart the IOC. This can be done either by running start_ibex_server but it is more efficient to console to the IOC and restart it using ctrl+x.
If you want to run an IOC in a mini-inst then:
- Add the device name to the "startup.txt" (
C:\Instrument\Settings\config\NDX<instrument>\configurations\startup.txt)- e.g. JSCO4180_01
- Then set macros in "globals.txt" (
C:\Instrument\Settings\config\NDX<instrument>\configurations\globals.txt)- e.g. JSCO4180_01__PORT=COM13
- Run
make iocstartupsin the epics terminal. This adds the device to the allowed running devices. - Start IBEX server as normal
start_ibex_server.bat - If you are running the IOC to allow a Vi in SECI to communicate then configure that vi (usually the pv prefix)
- e.g. Run
C:\LabVIEW Modules\Drivers\Jasco PU-4180 HPLC Pump\Jasco HPLC Pump - System.llb- Initialisation File change IOC Root toIN:CRISP:JSCO4180_01:
- e.g. Run
- Run the Vi