Been trying to set to zero the sensor as it has hovered around 2900ppb in a clean environment. I have been trying to set to Zero, however, everytime I'm clicking the Set Zero button in the DSDK Tool 0.42, it always returns ADC Range error.
Any way to solve this?
The text was updated successfully, but these errors were encountered:
The range error is usually caused by the device still reading too high. I
recommend that it is continuously powered for at least 12 hours. I've also
seen some users with unusually high indoor values. Try taking the device
outside for 15 minutes and zeroing there.