List of all settings in the program:
- He3PumpMaxVoltage Maximum voltage setting for the He3 pump
- He4PumpMaxVoltage "
- He4SwitchMaxVoltage "
- He3SwitchMaxVoltage "
- ControllerHe3HeaterVoltage, Maximum voltage the controller is allowed to use when heating the He3 pump.
- ControllerHe3SwitchVoltage, Maximum voltage the controller is allowed to use when heating the He3 switch.
- ControllerHe4SwitchVoltage, etc.
- ControllerHe4HeaterVoltage, etc.
- ControllerHe7StartTemperature, Temperature where the controller starts the control loop for heating the He3 and He4 pumps
- ControllerHeaterTemperatureSetpoint, Pump heater setpoint, Pumps are heated to this temperature while recycling or cooling down to 4K.
- ControllerHeatSwitchOnTemperature, Temperature where heat switches are assumed to be in the conducting state.
- ControllerHeatSwitchSafeValue, If the 2K plate is above this temperature, the heat switch heaters are turned off, since absorbing is useless.
- ControllerHeatupTemperature, Temperature setpoint used in the Heat up sequence.
- ControllerHe4StartTemperature, Temperature at which He4 evaporation is started. All He4 should be condensed at this temperature
- ControllerHe3StartTemperature, Temperature at which He3 evaporation is started, All He3 should be condensed at this temperature.
- ControllerDisableHeaterHeatSwitchTemperature, Pump heaters are disabled when heat switches are over this temperature, since the heat pumps are then conducting to the 2k plate.