@jast1982 jast1982 commented Jun 3, 2020

created a virtual inhalation sensor out of pressure sensor input
A sensor value > threshold detects a patients breath and directly allows
to go from exhalation to inhalation. Timing for subsequent breaths is
adjusted put pulled back 20% per breath if patient is not triggering
another breath

  • max pressure is taken from parameters

created a virtual inhalation sensor out of pressure sensor input
A sensor value > threshold detects a patients breath and directly allows
to go from exhalation to inhalation. Timing for subsequent breaths is
adjusted put pulled back 20% per breath if patient is not triggering
another breath
- max pressure is taken from parameters